Abstract

A liquid chromatographic–tandem mass spectrometry (LC–MS/MS) method for the determination of taselisib (GDC-0032, RO5537381) concentrations in human plasma has been developed and validated to support bioanalysis of clinical samples. Solid phase extraction (SPE) was used to extract plasma samples (50μL) and the resulting samples were analyzed using reversed phase chromatography and mass spectrometry coupled with an atmospheric pressure chemical ionization interface. The mass analysis of taselisib was performed using multiple reaction monitoring transitions in positive ionization mode. The method was validated over the calibration curve range 0.400–400ng/mL using linear regression and 1/x2 weighting. The within-run relative standard deviation (%RSD) ranged from 1.3 to 5.6%, while the between-run %RSD varied from 2.0 to 4.5% for LLOQ, low, medium, medium high and high QCs. The accuracy ranged from 94.7 to 100.3% of nominal for within-run and 96.0–99.0% of nominal for between-run for the same QCs. Extraction recovery of taselisib was between 83.8% and 92.9%. Stability of taselisib was established in human plasma for 977days at −20°C and −70°C and established in sample extracts for 96h when stored at 2 − 8°C. Stable-labeled internal standard was used to minimize matrix effects. Mean single dose pharmacokinetic parameters determined using this method for a phase I/II clinical trial were: Cmax=35.2ng/mL, AUC0−inf=1570ngh/mL, and T1/2=39.3h.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.