Abstract

This work discusses the strengths, limitations and validity of a novel arterial spin labeling technique when used specifically to measure perfusion in limb skeletal muscle. The technique, flow-driven arterial water stimulation with elimination of tissue signal (FAWSETS), offers several advantages over existing arterial spin labeling techniques. The primary goal of this study was to determine the perfusion signal response to changes in net hind limb flow that were independently verifiable. The range of perfusate flow was relevant to skeletal muscle during mild to moderate exercise. Localized, single voxel measurements were acquired from a 5 mm-thick slice in the isolated perfused rat hind limb at variable net flow rates. The results show that the perfusion signal is linearly proportional to net hind limb flow with a correlation coefficient of 0.974 (p = 0.0013). FAWSETS is especially well suited for studies of skeletal muscle perfusion, where it eliminates the need to compensate for magnetization transfer and arterial transit time effects. A conceptual discussion of the basic principles underlying these advantages is presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.