Abstract

Measurement of nanometre-sized aerosol particles is based on particle number concentration measurements. The commonly used method for providing traceability for these measurements involves charging and electrical counting of aerosol particles. This method requires that the particles are singly charged or that the average charge is exactly known, neither of which is easy to ensure. In the device called a single charged aerosol reference (SCAR), the fraction of multiply charged particles is minimal due to the novel operating principle of electrical charging and subsequent growth. In this study the SCAR was validated as a primary particle number concentration standard. The average charge of the output aerosol was evaluated for the whole operational particle size range. For this, the effect of the size distribution of the primary nanoaerosol and the output number concentration on the fraction of doubly charged and neutral particles was measured. It was found that the uncertainty caused by assuming singly charged particles is only 0.16%. A full uncertainty analysis was carried out for a condensation particle counter (CPC) calibration. According to the results, the relative expanded uncertainty of calibration was 3.0%. This represents a typical uncertainty level achieved in CPC calibrations performed with SCAR. As a result of this study, SCAR was validated as a particle number concentration standard suitable for traceable calibration of particle counting instruments in the particle size range from 10 nm to 500 nm.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call