Abstract

With the aid of indocyanine green (ICG), lymphatic architecture and function in both mice and humans has been successfully imaged non-invasively using near-infrared (NIR) fluorescence imaging devices. Maximal measurement sensitivity of NIR fluorescence imaging devices is needed for "first-in-humans" molecularly targeting NIR fluorescence agents that are brighter than non-specific ICG. In this study, we developed a solid phantom and measurement approach for the quantification of excitation light leakage and measurement sensitivity of NIR fluorescence imaging devices. The constructed solid phantom, consisting of quantum dots impregnated onto specularly reflective surface, shows long-term stability and can be used as a traceable fluorescence standard. With the constructed solid phantom, the intensified CCD (ICCD)-based device demonstrated more than 300% higher measurement sensitivity compared to the Electron Multiplying CCD (EMCCD) based device when integration time was maintained less than 1.0 s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.