Abstract

Two simple and sensitive spectrofluorometric methods have been developed and validated for determination of amlodipine besylate (AML) in tablets. The first method was based on the condensation reaction of AML with ninhydrin and phenylacetaldehyde in buffered medium (pH 7.0) resulting in formation of a green fluorescent product, which exhibits excitation and emission maxima at 375 and 480 nm, respectively. The second method was based on the reaction of AML with 7-chloro-4-nitro-2,1,3-benzoxadiazole (NBD-Cl) in a buffered medium (pH 8.6) resulting in formation of a highly fluorescent product, which was measured fluorometrically at 535 nm ( λ ex, 480 nm). The factors affecting the reactions were studied and optimized. Under the optimum reaction conditions, linear relationships with good correlation coefficients (0.9949–0.9997) were found between the fluorescence intensity and the concentrations of AML in the concentration range of 0.35–1.8 and 0.55–3.0 μg ml −1 for ninhydrin and NBD-Cl methods, respectively. The limits of assays detection were 0.09 and 0.16 μg ml −1 for the first and second method, respectively. The precisions of the methods were satisfactory; the relative standard deviations were ranged from 1.69 to 1.98%. The proposed methods were successfully applied to the analysis of AML in pure and pharmaceutical dosage forms with good accuracy; the recovery percentages ranged from 100.4–100.8 ± 1.70–2.32%. The results were compared favorably with those of the reported method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call