Abstract

This study presents a selective high-performance liquid chromatography (HPLC) with diode array detection (DAD) method for the simultaneous estimation of diclofenac sodium and lidocaine hydrochloride in presence of four of their related substances and potential impurities, namely, 2,6-dimethylaniline (DMA), 2,6-dichloroaniline (DCA), N-phenyl-2,6-dichloroaniline (PDCA), and N-chloroacetyl-N-phenyl-2,6-dichloroaniline (CPDCA). Some of these related substances are reported as degradation products as well. Effective chromatographic separation was achieved using Waters Symmetry C18 column, (3.9 × 150 mm, 5 μm particle size) with gradient elution of the mobile phase composed of 0.05 M orthophosphoric acid and acetonitrile. The gradient elution started with 5% (by volume) acetonitrile, ramped up linearly to 65% in 5 min then kept constant till the end of the run. The mobile phase was pumped at a flow rate of 1.5 mL min−1. The multiple wavelength detector was set at 220 nm, and quantification of both drugs was ba...