Abstract

A simple, accurate and sensitive reversed-phase high-performance liquid chromatographic (RP-HPLC) method has been developed for the estimation of ethyl 2-aminobenzoate (EAB) in a matrix type monolithic polymeric device and validated as per the International Conference on Harmonization guidelines. The analysis was performed isocratically on a ZORBAX Eclipse plus C18 analytical column (250 × 4.4 mm, 5 μm) and a diode array detector (DAD) using acetonitrile and water (75:25 v/v) as the mobile phase by keeping the flow-rate constant at 1.0 mL/min. Determination of EAB was not interfered in the presence of excipients. Inter- and intra-day relative standard deviations were not higher than 2%. Mean recovery was between 98.7 and 101.3%. Calibration curve was linear in the concentration range of 0.5-10 µg/mL. Limits of detection and quantification were 0.19 and 0.60 µg/mL, respectively. Thus, the present report put forward a novel method for the estimation of EAB, an emerging insect repellent, by using RP-HPLC technique.