Abstract

An accurate, sensitive and precise HPLC method was developed and validated for the routine analysis of pioglitazone hydrochloride in dosage forms. The analyte was chromatographed on a C-18 column using a mixture of acetonitrile and ammonium acetate buffer (pH 5.0) as mobile phase in ratio 60: 40 (v/v) at flow rate 1.0 mL/min. Linearity range was found to be 10–100 μg/mL with a correlation coefficient r2 = 0.9984. The method was validated for precision, accuracy, robustness, specificity and sensitivity, using bulk drug samples. Application of method in assay of bulk drug and tablets revealed mean recoveries range from 99.88–100.32%. Due to its simplicity, rapidity, high precision and accuracy, the proposed method may be used for determining pioglitazone hydrochloride in bulk and dosage forms.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.