Abstract
A highly sensitive and fully validated method was developed for the quantification of baclofen in human plasma. After adjusting the pH of the plasma samples using a phosphate buffer solution (pH 4), baclofen was purified using mixed mode (C8/cation exchange) solid-phase extraction (SPE) cartridges. Endogenous water-soluble compounds and lipids were removed from the cartridges before the samples were eluted and concentrated. The samples were analyzed using triple-quadrupole liquid chromatography-tandem mass spectrometry (LC-MS-MS) with triggered dynamic multiple reaction monitoring mode for simultaneous quantification and confirmation. The assay was linear from 25 to 1,000 ng/mL (r(2) > 0.999; n = 6). Intraday (n = 6) and interday (n = 15) imprecisions (% relative standard deviation) were <5%, and the average recovery was 30%. The limit of detection of the method was 5 ng/mL, and the limit of quantification was 25 ng/mL. Plasma samples from healthy male volunteers (n = 9, median age: 22) given two single oral doses of baclofen (10 and 60 mg) on nonconsecutive days were analyzed to demonstrate method applicability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.