Abstract
In the presented work identification and quantification of some anti-inflammatory and analgesic drugs, namely aceclofenac, diclofenac, paracetamol and para-aminophenol were carried out by validated HPLC-UV method. The chromatographic separation was achieved on HPLC C18 (100.0 × 2.1 mm, 1.7μm) column using isocratic mobile phase consisting of acetonitrile-phosphate buffer (50: 50, v/v) at a flow rate of 1.0 mL/min. The UV detection was carried out at 275 nm for all the compounds. The elution of aceclofenac, diclofenac, paracetamol and para-aminophenol was occurred at 7.0, 9.2, 2.0 and 4.2 min, respectively. The calibration curves were linear over the concentration range of 1–1000 μg/mL for aceclofenac and paracetamol, and 1–100 μg/mL for diclofenac and para-aminophenol. The developed method was validated according to ICH guidelines. The method was applied in the identification and quantitative determination of these compounds during routine quality control analysis and in stability studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.