Abstract

The natural methyl donor group, S-adenosylmethionine and its product, S-adenosylhomocysteine play an important role in many biochemical reactions involving transmethylation reactions. These compounds can be used as biomarkers in incipient diagnosis of various pathological disorders therefore the validation of a suitable method to routinely analysis of these compounds is very important. In this paper, a high performance liquid chromatrography method for S-adenosylmethionine and S-adenosylhomocysteine measurement as fluorescent 1,N(6)-ethanoderivatives from biological samples was validated in terms of selectivity, linearity range of the response (R > 0.9993), detection limit (9 × 10(-9) and 4.4 × 10(-9) mol L(-1)), the limit of quantitation (9.7 × 10(-9) and 5.7 × 10(-9) mol L(-1)), precision, trueness and robustness. The method for quantification simultaneous of these compounds is rapid, sensitive and precise and appropriate for clinical analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.