Abstract

A sensitive and specific GC/MS method for the determination of clozapine (CLZ) and its major metabolite norclozapine (NCLZ), in plasma has been developed, optimized and validated. Specimen preparation includes solid-phase extraction of both analytes using Bond-Elut Certify cartridge and further derivatization with TFAA. Clozapine-d8 was used as internal standard for the determination of CLZ and NCLZ. Limits of detection were 0.45 ng/mL for CLZ and 1.59 ng/mL for NCLZ, while limits of quantification were 1.37 ng/mL for CLZ and 4.8 ng/mL for NCLZ, as calculated by the calibration curves. The calibration curves were linear up to 600 ng/mL for CLZ and NCLZ. Absolute recovery ranged from 82.22% to 95.35% for both analytes. Intra- and interday accuracy was less than 7.13% and −12.52%, respectively, while intra- and interday precision was between 9.47% and 12.07%, respectively, for CLZ and NCLZ. The method covers all therapeutic range and proved suitable for the determination of CLZ and NCLZ not only in psychiatric patients but also in forensic cases with clozapine implication.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.