Abstract
Objective: Tocodynamometry is the most common method of labor evaluation but most clinicians would agree it has limited utility before 26 weeks of gestation. The obesity epidemic has further reduced our ability to accurately detect uterine contractions using the tocodynamometer at any gestational age. We sought to design and test a novel contraction monitor that bypasses the maternal abdomen. Methods: An optimized version of an intravaginal electrohysterographic ring device was tested in an ovine model. The device and its methodology as well as the tocodynamometer were validated against the current gold standard uterine activity monitor, the intrauterine pressure catheter in six sheep at varying gestational ages. Results: Both the intravaginal ring device and the tocodynamometer correlated well with IUPC, r = 0.69 and 0.73, respectively (p < 0.001). The number of contractions detected by each monitor remained similar even after accounting for confounders. Conclusions: These results suggest that uterine activity can be monitored from the vaginal interface in an ovine model and offers an alternative clinical tool for the detection of contractions in situations, in which tocodynamometry would be ineffective or intrauterine monitoring inappropriate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.