Abstract

A miniature argon arc has been designed and tested as a new transfer standard of spectral radiance for the wavelength range from 114 nm to 330 nm. Calibration has been performed using two primary standard sources: the hydrogen arc from 130 nm to 330 nm and the blackbody line radiator from 114 nm to 130 nm. The mini-arc provides an intense, stable, and reproducible uv continuum with dc power requirements of less than 1.5 kW. The arc characteristics have been investigated, and the sensitivity of the radiant power output to various operating parameters has been measured. The rms uncertainty in the spectral radiance is estimated to be 5.3% above 140 nm and 10.1% between 114 nm and 140 nm, due primarily to uncertainties in the primary standard sources.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.