Abstract

A mass spectrometer in which atoms from a 0.1-mm2 surface of a target are sputtered and ionized under bombardment by a beam of inert-gas ions has been perfected. To exploit the capabilities of this analytical tool for the detection of trace elements in solids, a drastic reduction of the instrumental background had to be achieved. This was accomplished by using spectroscopically pure structural material in the ion source and by eliminating all sources of hydrocarbons. In particular, since the presence of pump-oil contamination on the surface under analysis was readily detectable, the resulting cracking pattern obscured a large portion of the atomic-mass scale. A combination of liquid-nitrogen-trapped mercury diffusion pumps, zeolite sorption pumps, and sputter-ion pumps has reduced the instrumental background sufficiently to facilitate the routine parts-per-billion trace analysis of the surface and of the bulk of conducting and nonconducting solids.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.