Abstract

Vacuum technology is an integral part of any accelerator system. At IUAC we have a 15UD PELLETRON, superconduting LINAC, Low Energy Ion beam Facility and a 1.7MV pelletron. Vacuum requirement in these accelerators is ~10−8 torr. Various types of Vacuum pump are used in different zones of the accelerators depending on load. Since the whole accelerator is quite long, distributed pumps are placed in different sections as per load. In ion sources displacement type pump viz turbo-pumps are usually used as the gas load is quite high. In other parts of the accelerator combination of getter and ion pumps are used. It is very much necessary to isolate different sections for maintenance purpose. Proper valves are used to isolate the sections and to avoid vacuum accidents proper interlock system is introduced. If air goes in some sections accidentally, valves will close automatically to protect other sections. The talk will cover different types of pumps and interlock used in accelerators at IUAC.