Abstract

Some Bayard–Alpert gauges mounted on crotch chambers in the SPring-8 storage ring have indicated negative pressures at stored electron beam currents more than 5 mA. Simple measurements have shown that the negative pressure indications of these vacuum gauges are caused by influx of photoelectrons to their collectors. In order to measure the pressure in such a vacuum environment, the authors propose a vacuum gauge which self-compensates for photoelectrons from the external vacuum environment. The gauge has an added correcting electrode which detects the photoelectron current, only from the external environment, and then compensates to show the real pressure from the measured ion current of the gauge collector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.