Abstract
Pirani gauges are used in an ac bridge circuit to sense deviation from the set point, and with the aid of a porportional leak valve, control the pressure of laboratory-scale vacuum systems over the range of 0.1 to 10 000 μ of Hg. Pressure fluctuations from the controller action are negligible and stability of ±1% has been obtained over long periods without operator attention. Set point is easily adjustable by panel control knob. Commercial components are used except for the leak valve which is of special construction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.