Abstract
Chitosan-based sensor thin films were fabricated to detect trace amounts of lead ion using surface plasmon resonance (SPR) optical sensor. The gold surface used for SPR measure- ments was modified with chitosan and chitosan-tert-butylcalix[4]arene-tetrakis(N,N-dimethylthioacetamide) (chitosan-BCAT). Both chitosan and chitosan-BCAT layers were deposited on the gold surface by spin coating technique. The experiment has been carried out to monitor the SPR signals for lead ion with sensitive enhancement by chitosan and chitosan-BCAT layers. For both layers, the change in resonance angle (Δθ) is directly proportional to the concentration of lead ion solution. The higher amounts of Δθ were obtained for chitosan-BCAT film due to a specific binding of BCAT with lead ion. The chitosan-BCAT film enhanced the sensitivity of detection down to 0.03 ppm. Data analysis also has been done by Matlab software using Fresnel formula for multilayer system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.