Abstract

The potential use of activated carbon (AC) as an inexpensive and effective alternative sorbent material in thermal desorption is presented and validated for the analysis of aromatic volatile organic compounds (VOCs) such as benzene, toluene, m-xylene, and styrene (BTXS) in air. The optimum desorption conditions of an AC sampling tube (2 mg AC bed) were determined and compared with a commercial three-bed (Carbopack; C + B + X) tube sampler as a reference. The AC sampler exhibited good linearity (R2 > 0.99) and reproducibility (RSE of 2.38 ± 0.21%) for BTXS analysis. The AC tube sampler showed good storability (up to 3 d) and excellent recyclability (up to 50 cycles). An analysis of BTXS in ambient air showed excellent agreement between AC and CBX (bias < 5%). The 1% breakthrough volume values for 2 mg AC, when tested at 100 ppb of benzene as a sole component or in a BTXS mixture, were 10,000 or 5000 L g−1, respectively. The results of this study support the performance of AC as a suitable medium for sampling VOCs as reliable as high-cost commercial sorbent products.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.