Abstract

Follow-up examinations after flow diverter (FD) treatment for cerebral aneurysms typically involve magnetic resonance imaging (MRI) or digital subtraction angiography (DSA). However, MRI is prone to vascular defects due to metal artifacts from FD, and DSA carries a risk of ischemic complications. In the context of computed tomography angiography (CTA), this study compares the efficacy of ultra-high-resolution CT (UHRCT) and novel reconstruction techniques, such as model-based iterative reconstruction (MBIR), against conventional methods such as filtered back projection (FBP) and hybrid iterative reconstruction (IR), to determine if they are a viable alternative to DSA in clinical settings. A phantom study was conducted with the full-width half-maximum considered as the FD thickness. This study compared three reconstruction methods: MBIR, FBP, and hybrid IR. A clinical study was also conducted with 21 patients who underwent follow-up CTA after FD treatment. The FD's visibility was assessed using a 4-point scale in FBP, hybrid IR, and MBIR compared to cone-beam CT (CBCT) with angiographic systems. In the phantom study, FBP, hybrid IR, and MBIR visualized thinner FD thicknesses and improved detail rendering in that order. MBIR proved to be significantly superior in both the phantom and clinical study. UHRCT with MBIR is highly effective for follow-up evaluations after FD treatment and may become the first-choice modality in the future.