Abstract

The United States Pharmacopoeia dissolution apparatus 3 (reciprocating cylinder) was evaluated with respect to effects of changes in instrument parameters on drug release rate from six hydrophilic matrix formulations and one coated-bead formulation. Reciprocation rate had the largest effect on time to 50% release for all matrix formulations. Top mesh size had an effect on release rate for those formulations having an erosion component to the drug release mechanism. Bottom mesh size had no effect on release rate. For the coated-bead formulation, no parameter had an effect on release rate. In an assessment of the hydrodynamics of the reciprocating cylinders, conditions equivalent to the 50rpm paddle and 100rpm basket were determined to be at the extreme low end of the available reciprocation rate range.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.