Abstract

Many in-vitro experiments performed to study the response of thiol-containing proteins to changes in environmental redox potentials use dithiothreitol (DTT) to maintain a preset redox environment throughout the experiments. However, the gradual oxidation of DTT during the course of the experiments, and the interaction between DTT and other components in the system, can significantly alter the initial redox potential and complicate data interpretation. Having an internal reporter of the actual redox potential of the assayed sample facilitates direct correlation of biochemical findings with experimental redox status. Reversed-phase high-performance liquid chromatography (RP-HPLC) is a widely used, well-established tool for analysis and purification of biomolecules, including proteins and peptides. Here, we describe a simple, robust, and quantitative RP-HPLC method we developed and tested for determination of the experimental redox potential of an in-vitro sample at the time of the experiment. It exploits the specific UV-absorbance of the oxidized intrinsic DTT in the samples and retains the high resolving power and high sensitivity of RP-HPLC with UV detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call