Abstract
A CNN based method for cardiac MRI tag tracking was developed and validated. A synthetic data simulator was created to generate large amounts of training data using natural images, a Bloch equation simulation, a broad range of tissue properties, and programmed ground-truth motion. The method was validated using both an analytical deforming cardiac phantom and in vivo data with manually tracked reference motion paths. In the analytical phantom, error was investigated relative to SNR, and accurate results were seen for SNR>10 (displacement error <0.3mm). Excellent agreement was seen in vivo for tag locations (mean displacement difference = -0.02pixels, 95% CI [-0.73, 0.69]) and calculated cardiac circumferential strain (mean difference = 0.006, 95% CI [-0.012, 0.024]). Automated tag tracking with a CNN trained on synthetic data is both accurate and precise.
Accepted Version (Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.