Abstract

This study evaluated the dependence of the signal characteristics of time-spatial labeling inversion pulse (time-SLIP) on flow velocity and tag thickness to depict the pancreatic juice flow by analyzing signal profile using a tube phantom study. The tag edge property was evaluated by edge rise distance (ERD). For various slow flow velocities and tag thicknesses, the signal profile characteristics were evaluated using two indices: the tag center value (RTCV) reduction rate and the total signal value along the tube (TSVT). ERD, which was about 10% of the tag thickness, was higher for thicker tags, making slow flow detection difficult. TSVT was proportional to the thickness of the tag and was preserved irrespective of the flow velocity. RTCV became lower with higher flow velocity and decreased significantly with thinner tags. These results suggest that the visualization of pancreatic juice flow might improve stability by considering the appropriate tag thickness.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.