Abstract

The assessment procedure currently used for the accreditation of building acoustics laboratories is given in the standard ISO 140-2. It is based on the deviations between values obtained in the new laboratory and round robin results whereupon the number of third-octave values outside of the critical differences is simply counted. A new procedure is proposed which additionally accounts for the amount of deviation from the round robin mean value. Thus, it can be considered as a refined continuously scaled version of the current procedure. The new method has been checked by applying it to the data from a round robin survey of a heavy test wall. Besides the comparison between individual laboratory results and round robin mean results, an additional single measurement, carried out in each laboratory, is proposed to verify the results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.