Abstract

Proton-transfer-reaction mass spectrometry (PTR-MS) is a useful tool in ambient trace gas analysis, especially for the analysis of oxygenated volatile organic compounds (OVOC). Many OVOCs are produced during photooxidation of volatile organic compounds and contribute to both the gas phase and secondary organic aerosols (SOA). The inlet system of the PTR-MS instrument was modified to allow also for the measurement of the particulate phase of an aerosol with a high time resolution. The new inlet consists mainly of a denuder to strip off the gas phase, and a heater (120/150 degrees C) to vaporize the aerosol particles. This inlet system was tested with pinonic acid particles generated with a nebulizer and SOA particles formed during the photooxidation of 1,3,5-trimethylbenzene and alpha-pinene with NO(x) in a smog chamber. The performance of this new technique is discussed and the partitioning coefficients for the oxidation products are estimated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.