This study was registered in ClinicalTrials.gov (NCT04202237). The Moor FLPI-2 blood flow imager was used in 4 patients (1 Simultaneous Pancreas and Kidney, 2 deceased, and 1 living donor kidney transplants) during reperfusion to capture reperfusion data. The following parameters were measured: flux (average speed × concentration of moving red blood cells in the sample volume), doppler centroid, total and valid pixels, valid rate, and total and valid area. Flux data were analyzed with Moor FLPI analysis software. The perfusion characteristics and flux images correlated with initial graft function. LSCI is a safe, noncontact imaging modality that provides real-time, accurate, high-resolution, full field blood flow images and a wide range of flux data to objectively quantify organ reperfusion intraoperatively in kidney/kidney-pancreas transplantation. This modality could be used to develop a robust numerical quantification system for the evaluation and reporting of intraoperative organ perfusion, and aid intraoperative decision-making. Perfusion data could be combined with biomarkers and immunological parameters to more accurately predict graft outcomes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.