Abstract

The cochlear implant (CI) fitting level prediction accuracy of electrically-evoked compound action potential (ECAP) should be enhanced by the addition of demographic data in models. No accurate automated fitting of CI based on ECAP has yet been proposed. We recorded ECAP in 45 adults who had been using MED-EL CIs for more than 11 months and collected the most comfortable loudness level (MCL) used for CI fitting (prog-MCL), perception thresholds (meas-THR), and MCL (meas-MCL) measured with the stimulation used for ECAP recording. Linear mixed models taking into account cochlear site factors were computed to explain prog-MCL, meas-MCL, and meas-THR. Cochlear region and ECAP threshold were predictors of the three levels. In addition, significant predictors were the ECAP amplitude for the prog-MCL and the duration of deafness for the prog-MCL and the meas-THR. Estimations were more accurate for the meas-THR, then the meas-MCL, and finally the prog-MCL. These results show that 1) ECAP thresholds are more closely related to perception threshold than to comfort level, 2) predictions are more accurate when the inter-subject and cochlear regions variations are considered, and 3) differences between the stimulations used for ECAP recording and for CI fitting make it difficult to accurately predict the prog-MCL from the ECAP recording. Predicted prog-MCL could be used as bases for fitting but should be used with care to avoid any uncomfortable or painful stimulation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call