Abstract

We use computational fluid dynamics to compare the shear rate and turbulence in an advanced in vitro gastric model (TIMagc) during its simulation of fasted state Migrating Motor Complex phases I and II, with the United States Pharmacopeia paddle dissolution apparatus II (USPII). A specific focus is placed on how shear rate in these apparatus affects erosion-based solid oral dosage forms. The study finds that tablet surface shear rates in TIMagc are strongly time dependant and fluctuate between 0.001 and 360 s−1. In USPII, tablet surface shear rates are approximately constant for a given paddle speed and increase linearly from 9 s−1 to 36 s−1 as the paddle speed is increased from 25 to 100 rpm. A strong linear relationship is observed between tablet surface shear rate and tablet erosion rate in USPII, whereas TIMagc shows highly variable behavior. The flow regimes present in each apparatus are compared to in vivo predictions using Reynolds number analysis. Reynolds numbers for flow in TIMagc lie predominantly within the predicted in vivo bounds (0.01-30), whereas Reynolds numbers for flow in USPII lie above the predicted upper bound when operating with paddle speeds as low as 25 rpm (33).