Abstract

BackgroundBlood collected in conventional EDTA tubes requires laboratory analysis within 48 hours to provide valid CD4 cell count results. This restricts access to HIV care for patients from rural areas in resource-constraint settings due to sample transportation problems. Stabilization Tubes with extended storage duration have been developed but not yet evaluated comprehensively.ObjectiveTo investigate stability of absolute CD4 cell count measurement of samples in BD Vacutainer CD4 Stabilization Tubes over the course of 30 days.MethodsThis was a laboratory-based method comparison study conducted at a rural district hospital in Beitbridge, Zimbabwe. Whole peripheral blood from 88 HIV positive adults was drawn into BD Vacutainer CD4 Stabilization Tubes and re-tested 1, 2, 3, 5, 7, 14 and 30 days after collection on BD FacsCount and Partec Cyflow cytometers in parallel. Absolute CD4 cell levels were compared to results from paired samples in EDTA tubes analysed on BD FacsCount at the day of sample collection (references methodology). Bland-Altman analysis based on ratios of the median CD4 counts was used, with acceptable variation ranges for Limits of Agreements of +/-20%.ResultsDifferences in ratios of the medians remained below 10% until day 21 on BD FacsCount and until day 5 on Partec Cyflow. Variations of Limits of Agreement were beyond 20% after day 1 on both cytometers. Specimen quality decreased steadily after day 5, with only 68% and 40% of samples yielding results on BD FacsCount and Partec Cyflow at day 21, respectively.ConclusionsWe do not recommend the use of BD Vacutainer CD4 Stabilization Tubes for absolute CD4 cell count measurement on BD FacsCount or Partec Cyflow due to large variation of results and decay of specimen quality. Alternative technologies for enhanced CD4 testing in settings with limited laboratory and sample transportation capacity still need to be developed.

Highlights

  • For HIV infected persons, in particular in resource-limited settings, treatment decisions are largely contingent on absolute CD4+ T-lymphocyte cell counts in venous blood

  • Absolute CD4 cell levels were compared to results from paired samples in Ethylenediaminetetraacetic acid (EDTA) tubes analysed on BD FacsCount at the day of sample collection

  • Differences in ratios of the medians remained below 10% until day 21 on BD FacsCount and until day 5 on Partec Cyflow

Read more

Summary

Background

Blood collected in conventional EDTA tubes requires laboratory analysis within 48 hours to provide valid CD4 cell count results. This restricts access to HIV care for patients from rural areas in resource-constraint settings due to sample transportation problems. Stabilization Tubes with extended storage duration have been developed but not yet evaluated comprehensively

Objective
Methods
Results
Conclusions
Introduction
Material and Methods
Discussion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.