This study was performed to clarify the differences in blood flow strength, blood vessel diameter, and post-labeling delay (PLD) by physical experiments, and to examine whether bright vessel appearance (BVA) can be observed by arterial spin labeling (ASL). We introduced simulated blood flow (25 cm/sec, 12.5 cm/sec) using a specially made phantom of fixed tubes in a plastic container. At each speed, we scanned at several points of PLD using ASL imaging. We measured the signal in the tube to obtain a signal intensity (SI). We revised the T1 level from the SI and obtained SIblood. We used SItissue with normal perfusion measured from obtained clinical images by ASL and compared it with SIblood. In tubes with a narrow inner diameter, the signal slightly decreased. SI also decreased under slow flow compared with fast flow. At each flow rate, SIblood significantly exceeded SItissue. PLD distinguishes spin in brain tissue from 1525 msec to 2525 msec, and it can be observed. As spin signal decreases when the flow rate is slow, attention is necessary for observation. Assessment at PLD1525-2525 msec where normal perfusion was obtained suggested that BVA can be observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.