Abstract

Conclusions: EABR is a reliable and effective way of objectively confirming device function and implant-responsiveness of the peripheral auditory neurons up to the level of the brainstem in cases of inner ear malformation. Objective: To investigate the usefulness of measuring the intra-operative electrically evoked compound action potential (ECAP) and electrically evoked auditory brainstem response (EABR) in patients with and without congenital inner ear anomalies during cochlear implantation. Method: Thirty-eight consecutive children (40 ears) aged 5 or younger with congenital profound hearing loss. Twenty-four (25 ears) lacked congenital inner ear anomalies. The 14 patients (15 ears) with a malformation had common cavities (four ears), incomplete partition type I (three ears), cochlea hypoplasia type III (three ears), enlarged vestibular aqueduct (four ears), and cochlear nerve canal stenosis (one ear). Main outcome measures are ECAP and EABR responses. Results: Of the 25 ears lacking any malformation, 21, three, and one showed ‘Good’, ‘Variable’, and ‘No’ ECAP responses, respectively, and 24 and one showed ‘Good’ and ‘Variable’ intra-cochlear responses, respectively. Of the 15 ears with a malformation, two showed ‘Good’ ECAP responses, nine had ‘Variable’ ECAP responses, and four showed ‘No’ ECAP responses. Moreover, five showed ‘Good’ EABR responses and 10 showed ‘Variable’ EABR responses.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call