Abstract
Purpose : To determine the usefulness of for measuring the systemic circulation time. Material and Methods : Systemic circulation time, defined as elapsed time from the injection of solution to the point of pharyngeal burning sensation, was measured in 63 volunteers. was injected into a su-perficial vein of an upper extremity. Using dynamic electron beam computed tomography at the level of the abdominal aorta and celiac axis, a timeintensity curve was plotted, and for these two locations, maximal en-hancement time was compared. Results : For 60 of the 63 subjects, both systemic circulation time and maximal enhancement time were deter-mined. Average systemic circulation time was 17.4 (SD: 3.6) secs. and average maximal enhancement times at the level of the abdominal aorta and celiac axis were 17.5 (SD: 3.0) secs. and 18.5 (SD: 3.2)secs., respectively. Correlation coefficients between systemic circulation time and maximal enhancement time for the abdominal aorta and celiac axis were 0.73 (p can be used instead of a test bolus study.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.