Abstract

Respiratory-triggered-diffusion-weighted imaging (R-DWI) of the liver often results in poor image quality under the diaphragmatic dome on the cephalic side of the liver (hepatic dome) secondary to magnetic field inhomogeneity in liver magnetic resonance imaging (MRI). Hence, the usefulness of additional breath-hold-DWI (B-DWI) focusing on the hepatic dome was investigated. A total of 22 patients (14 men and 8 women; mean age 69.0±11.7 years) who underwent ethoxybenzyl (EOB)-MRI at our hospital between July and August, 2022 using a 3.0 T MRI system were included. One radiologist and three radiology technologists visually assessed the visibility of R-DWI and B-DWI in the hepatic dome on a 4-point scale (1 to 4). Additionally, the apparent diffusion coefficient (ADC) values of the hepatic parenchyma on each DWI were compared. B-DWI improved visibility in the hepatic dome compared to R-DWI (2.67±0.71 vs. 3.25±0.43, p<0.05). No significant difference was found in the ADC values for each DWI. B-DWI has excellent visibility in the hepatic dome and is expected to complement R-DWI. Therefore, B-DWI is very useful as an additional imaging in EOB-MRI.