Abstract

An X-ray diffraction procedure for quantitative chrysotile asbestos analysis in a gypsum-based matrix is discussed. The procedure includes calibration standards preparation, an EDTA treatment to remove gypsum, sample milling, and X-ray diffraction analysis. The effect of each step is discussed. Morphological and crystal structural changes due to the sample preparation procedure were also characterized using techniques such as SEM, IR, NMR, and ICP. Through a statistically designed experiment, it is demonstrated that X-ray diffraction can be used as a reliable technique in gypsum-based matrices to determine chrysotile concentration from 0.25% to 2.5% with a standard deviation of 0.07%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.