Abstract

A comparison was made of the performance of swing-out and vertical ultracentrifuge rotors in the detection of rubella-specific IgM on a sucrose density gradient. Tests were performed on 30 sera, of which 11 were found to contain rubella-specific IgM by both methods. The centrifugation time for the swing-out rotor was 16 hours at 35,000 rpm. This was reduced to 2 hours using the vertical rotor at 50,000 rpm. Routine use of the vertical rotor would allow a reduction in the time taken to perform the test, increase the number of sera tested each time, and reduce wear on the ultracentrifuge.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.