Abstract
Background/Aims: To evaluate the use of the NeuroMate stereotactic robot with a novel ultrasound registration system for movement disorder surgery (MDS). Methods: Using the robot in a frameless mode, 51 patients underwent MDS. Surgical planning was carried out using MRI data obtained more than 24 h before surgery. Results: 37 out of 50 targets in the subthalamic nucleus were satisfactorily identified with a single microelectrode trajectory and the final electrode positions were at a mean distance of 1.7 mm from the calculated target. There was a significant improvement in motor scores of the Unified Parkinson’s Disease Rating Scale III (off medication) at 6 (43%) and 18 months (51.7%) compared to pre-operative scores (p < 0.05). Conclusions: The frameless robot using only MRI data can be used for MDS. The temporal separation of imaging from the surgical procedure provides additional time for detailed image analysis and planning.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.