Abstract

ObjectiveTo investigate the clinical feasibility of the injection test for balloon placement during oesophageal pressure measurement in patients without spontaneous breathing.MethodsThe injection test was performed in 12 mechanically ventilated patients under deep sedation and paralysis. During withdrawal of the balloon from the stomach and air injection into the gastric lumen of the catheter, the presence of the injection test wave in the balloon pressure tracing indicated that the whole balloon was positioned above the lower oesophageal sphincter (LES). The positive pressure occlusion test was performed at different balloon positions.ResultsIn each patient, the injection test wave appeared at a distinct balloon depth, with a mean ± standard deviation of 41.9 ± 3.3 cm and range from 37 cm to 47 cm. The optimal ratio of changes in the balloon and airway pressure (0.8–1.2) during the positive pressure occlusion test was obtained when the balloon was located 5 cm and 10 cm above the LES in nine (75%) and three (25%) patients, respectively.ConclusionsThe injection test is feasible for identification of the whole balloon position above the LES during passive ventilation. The middle third of the oesophagus might be the optimal balloon position.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.