Abstract

Computing dissimilarity profiles between spectra within a chromatographic peak and comparing them to a base profile for a pure peak can be a sensitive method for detecting coelution. The power of this method is limited by noise and non-idealities. For high maximum absorbances, non-idealities prevail over noise. In current practice, either the maximum absorbances are limited, or the base profile is computed from injections of pure standards. The first approach leads to a loss in detection power, the second requires substantial extra work, since the standards have to be obtained and they have to be measured separately for each peak of interest in the chromatogram. Here, we show that an approximation to the base profile can be computed from realistic simulations of homogeneous data accounting for potential non-idealities. This is demonstrated using several graphical displays. As a result, higher maximum absorbances leading to better impurity detection can be used without extra experimentation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.