Abstract

In order to avoid the influence of pre-analytical steps, the following study was performed by using sterile blood spiked with defined loads of microorganisms as inoculum. Time-to-Detection (TTD) was evaluated for the most frequently encountered bacteria and fungi in septicemia, comparing two commercially available blood culture systems, BACTEC™ 9240 (Becton Dickinson) and BacT/ALERT™ (Organon Teknika). A specific medium, Bactec Mycosis™ IC/F (Becton Dickinson), was compared with the Bactec Plus Aerobic (Becton Dickinson) and FAN Aerobic (Organon Teknika) media for recovery of fungi in general and in case of mixed bacterial/fungal septicemia. The results show that the BACTEC™ system detects nearly all enrolled microorganisms significantly faster than the BacT/ALERT™; the anaerobic vial contributes to the detection of anaerobes and facultative anaerobes and, in the case of BACTEC™, shortens TTD; the Bactec Mycosis™ IC/F bottle shortens TTD of fungi.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.