Abstract
ObjectiveUsing a US Food and Drug Administration (FDA) emergency use authorization (EUA) reverse transcription polymerase chain reaction (RT-PCR) method, we examined the analytic performance accuracy of saliva specimens as compared to nasopharyngeal (NP) specimens in symptomatic patients. Correlation between test results and symptoms was also evaluated.MethodsOver a 5-week period in 2020, 89 matched saliva and nasopharyngeal swabs were collected from individuals exhibiting symptoms consistent with SARS-CoV-2. Specimens were tested with an FDA EUA-approved RT-PCR method, and performance characteristics were compared.ResultsThe concordance rate between saliva and nasopharyngeal testing was 93.26%. The mean cycle threshold value of saliva when compared to the NP specimen was 3.56 cycles higher. As compared to NP swab, saliva testing demonstrates acceptable agreement but lower sensitivity.ConclusionWhen compared to a reference method using NP swabs, the use of saliva testing proved to be a reliable method. Self-collected saliva testing for SARS-CoV-2 allows for a viable option when trained staff or collection materials are in short supply.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.