Abstract

En face electron fields to boost inguinal lymphatics have been used by oncologists for many years. With the introduction of multileaf collimators (MLC) and independent jaws, the practice of creating segmental fields to boost areas of interest has expanded. Typical anterior-posterior opposing field treatment of the pelvis may now be enhanced to include additional anterior segments to boost lymphatic tissue at a predetermined depth. This report illustrates the clinical implementation of one such segmental boost technique. Computer generated isodose plans utilize manual contour and CT-generated data for analysis of inguinal lymphatic depths. Potential areas of field overlap are discussed as well as the use of combined 6 and 15 MV photon energies to reduce areas of inhomogeneous dose. Technical details associated with MLC field size limits and other clinical factors are also discussed in relationship to smooth treatment delivery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.