Abstract

BackgroundAlthough many studies have evaluated the success of porous tantalum component in primary and revision cementless total knee arthroplasty (TKA) cases, few studies have reported the breakage of porous tantalum monoblock patella component (PTMP). The purpose of this study was to investigate the radiographic outcomes of breakage of PTMP at short-term follow up. MethodsThis was a retrospective study of 49 consecutive knees in 43 patients who underwent TKA surgery between September 2012 and March 2014. Bone marrow contents/tissue volumes (BMC/TV. mg/cm3) were evaluated using 2-dimensional osteomorphometry software with 2D- multi-detector-row computed tomography (2D-MDCT) to measure fixation of the bone-prosthesis interface with patella at 6, 12, and 24 months after surgery. Clinical follow up was obtained by reviewing each patient at orthopedic record. ResultsWe experienced a case that resulted in breakage of PTMP without traumatic episode at 26 months after surgery. Relative changes of Bone Mineral Contents per Tissue Volume. (BMC/TV mg/cm3) at ROI. 1 (superior border of the patella) and ROI. 2 (peg of patella) were significantly higher than ROI.3 (inferior border of patella) at 6, 12 and 24 months after surgery (p < 0.001. p < 0.01). Patient with breakage of PTMP has patella baja (Insall- Salvati ratio: 0.72), and the relative change of BMC/TV at ROI.1 and 2 for patients who experienced a breakage of PTMP was higher than any other patients. ConclusionThe present study revealed that the use of PTMP should be avoided for patients with patella baja.