Abstract

By utilizing flow-controlled PLL-g-PEG and PLL-g-PEGbiotin modification of predefined regions of a poly(dimethylsiloxane) (PDMS) micro-fluidic device, with an intentionally chosen large (approximately 1 cm2) internal surface area, we report rapid (10 min), highly localized (6 x 10(-6) cm2), and specific surface-based protein capture from a sample volume (100 microL) containing a low amount of protein (160 attomol in pure buffer and 400 attomol in serum). The design criteria for this surface modification were achieved using QCM-D (quartz crystal microbalance with energy dissipation monitoring) of serum protein adsorption onto PLL-g-PEG-modified oxidized PDMS. Equally good, or almost as good, results were obtained for oxidized SU-8, Topas, and poly(methyl metacrylate) (PMMA), demonstrating the generic potential of PLL-g-PEG for surface modification in various micro-fluidic applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.