Abstract

An inductively coupled plasma atomic emission spectroscopy method for speciation of Ag, Cu, Bi, Sb, As, Ni, Cd, Zn, Sn, and Fe in grade lead is described. The spectrometer was calibrated using water solutions with known concentrations of trace elements. The matrix effects were taken into account using an internal standard added to the reference-sample, calibrated-blank, reagent-blank, and tested-sample solutions. The metrological examination of the technique was accomplished. The repeatability rms error does not exceed 0.10. The systematic error is not statistically significant.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.