Abstract

Disaster victim identification (DVI) presents a number of physical and legal challenges, involving the degeneration of human remains and legal obstacles to forensic examinations. One non-invasive method for positive identification may be the use of a pacemaker programmer to detect and obtain data from pacemakers recovered from unidentified remains. To test the usefulness of this method, this investigation examined the efficiency and utility of 5 different pacemaker programmers in the positive identification of victims of the March 2011 tsunami in Japan at 8 disaster sites in May 2011. On scanning 148 sets of remains, data were successfully obtained from 1 implant in 1 set of remains, allowing for the rapid positive identification of the individual. Scanning pacemakers with pacemaker programmers can be a non-invasive method of positive identification that meets Japanese legal and institutional requirements, but this method is ineffective without a preceding whole-body X-ray scan.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.