Abstract

Transesophageal echocardiography (TEE) is a trusted method of sizing atrial septal defect (ASD) prior to percutaneous closure but is invasive, uncomfortable, and may carry a small risk of morbidity and mortality. Magnetic resonance imaging (MRI) may be useful non-invasive alternative in such patients who refuse or are unable to tolerate TEE and may provide additional information on the shape of the A0SD. To validate the accuracy of ASD sizing by MRI compared with TEE. Twelve patients (mean age 30 years; range 11-60 years) scheduled for ASD closure underwent TEE, cine balanced fast field echo MRI (bFFE-MRI) in four-chamber and sagittal views and phase-contrast MRI (PC-MRI) with reconstruction using the two orthogonal planes of T2-weighted images as planning. The average of the three longest measurements for all imaging modalities was calculated for each patient. Mean maximum ASD length on TEE was 18.8 +/- 4.6 mm, mean length by bFFE-MRI was 20.0 +/- 5.0 mm, and mean length by PC-MRI was 18.3 +/- 3.6 mm. The TEE measurement was significantly correlated with the bFFE-MRI and PC-MRI measurements (Pearson r = 0.69, p = 0.02 and r = 0.59, p = 0.04, respectively). The mean difference between TEE and bFFE-MRI measurements was -1.2mm (95% CI: -3.7, 1.3) and between TEE and PC-MRI was 0.5 mm (95% CI: -1.9, 2.9). Bland-Altman analysis also determined general agreement between both MRI methods and TEE. The ASDs were egg-shaped in two cases, circular in 1 patient and oval in the remaining patients. ASD sizing by MRI using bFFE and phase-contrast protocols correlated well with TEE estimations. PC-MRI provided additional information on ASD shapes and proximity to adjacent structures.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call