Abstract

AbstractA method of determining the absolute calibration of a gas‐flow orifice without the use of gas holders or any comparative device is described. The method is based on the application of the momentum balance, as well as the energy balance, to the flow of the gas. The application requires the measurement of pressures on the face of the orifice in addition to the usual pressure‐drop measurements along the axis of flow.Orifice coefficients determined by the force‐momentum principle are shown to agree within an average deviation of 1.4% with those determined by other standard techniques. Also the application of the force‐momentum principle demonstrates clearly why orifice coefficients are much less than unity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.