Abstract

On-capillary large volume sample stacking in capillary electrophoresis, with hydrodynamic injections of up to 300 s, can be used to significantly improve the LOD of selected cationic basic drugs and metal chelates by up to two orders of magnitude. This technique can be used to lower the LOD for a basic drug such as clenbuterol down to 3.2 × 10 - 8 mol dm - 3 for a 240 s injection from a value of 2.3 × 10 - 6 mol dm - 3 for conventional CZE using a 3 s hydrodynamic injection. Cobalt, as its cationic 5-Br-PADAP chelate, can be determined down to an LOD of 5.0 × 10 - 8 mol dm - 3 for a 300 s hydrodynamic injection as compared with an LOD of 5.0 × 10 - 7 mol dm - 3 for a 30 s hydrodynamic injection. This technique for such cations requires the use of the electroosmotic flow modifier cetyltrimethylammonium bromide for sample solvent removal during stacking. The cations are then determined at a UV/VIS detector after polarity switching.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.